Hampden Engineering Corporation MVO-1-PB
Ground Integrity Tester


Part NumberIncluded Options
Select Items To Quote
RENT/LEASE BUY USED
HAMD-MVO-1-PB 
Catalog Description

Hampden Engineering Corporation MVO-1-PB

HAMD-MVO-1-PB Ground Integrity Tester

The Hampden Model MVO-1-PB Ground Integrity Tester provides a simple and
convenient method of assuring the proper operation of hospital ground systems.

Operation
The Hampden Ground Integrity Tester may be operated in either of two modes:
ohms or millivolts. In the millivolts mode, it is used to measure the
difference in potential between the reference grounding point and exposed
conductive surfaces in a patient vicinity.
In the ohms mode, it is used to measure the resistance between the reference
grounding point and the ground contact of each receptacle in the patient
vicinity.

Features
o Measures both resistance and potential difference
o Color coded scale provides qualitative testing
o Meter is self-calibrating to internal standard
o Ohms scale is zeroed without shorting leads
LengthWidthHeightShip Wt
13.34 cm 17.53 cm 6.1 cm 3.64 kg
    