Home >> Products >> FLU-5522A
Other Configurations Available.
Call our experienced Sales Staff.
 
 
Fluke Mfg Co 5522A
Multi Product Calibrator


Part NumberIncluded Options
Select Items To Quote
RENT/LEASE BUY USED
FLU-5522A/PQ/SC1100/Power Quality Option/1.1GHz Scope Cal Option
FLU-5522A/SC1100/1.1GHz Scope Cal Option
Catalog Description

Fluke Mfg Co 5522A

FLU-5522A Multi Product Calibrator.


The FLU-5522A Multi-Product Calibrator addresses a wide calibration workload
and comes with internal and external protection features that protect it
against damage and make it easier to transport for on-site or mobile
calibration. The 5522A can also be fully automated with MET/CAL Plus
Calibration Management Software (version 7.3 SP1 or later). It is the ideal
calibrator for metrology professionals who need to perform resistance
calibration, voltage calibration, current calibration and more.

The FLU-5522A calibrator covers many of the electronic test tools including:
Handheld and bench meters (analog and digital) up to 6 digits
Current clamps and clamp meters, Thermocouples and RTDs,
Process calibrators, Data loggers, Strip and chart recorders, Watt meters
Power harmonics Analyser, Panel meters, Graphical multimeters
Power quality Analyser (with option)
Analog or digital handheld and bench oscilloscopes to 600 MHz or 1.1 GHz (with
option)

General Specifications:
Warmup Time: Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time: Less than 5 seconds for all functions and ranges except as
noted.

Standard Interfaces: IEEE-488 (GPIB), RS-232
Line Power: Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 Hz to 63 Hz
Line Voltage Variation: +/- 10 % about line voltage setting
Power Consumption: 600 VA

Temperature
Operating: 0 C to 50 C
Calibration (tcal): 15 C to 35 C
Temperature Coefficient: Temperature coefficient for temperatures outside
tcal +/-5 C is 10 % of the stated specification per C.

Relative Humidity
Operating: <80 % to 30 C, <70 % to 40 C, <40 % to 50 C

Safety: IEC 61010-1: Overvoltage CATII, Pollution Degree 2
Output Terminal Electrical Overload Protection: Provides reverse-power
protection, immediate output disconnection, and/or fuse protection on the
output terminals for all functions. This protection is for applied external
voltages up to +/-300 V peak.
Analog Low Isolation: 20 V normal operation, 400 V peak transient
Electromagnetic Environment : IEC 61326-1: Controlled
Electromagnetic Compatibility: If used in areas with Electromagnetic fields
of 1 to 3 V/m from 0.08-1GHz, resistance outputs have a floor adder of 0.508?.
Performance not specified above 3 V/m
LengthWidthHeightShip Wt
47.24 cm 43.18 cm 17.78 cm 35.91 kg
    
N/A Currently not available for sale

 